Uvod
The Frigga M2D is a reusable USB temperature data logger designed for accurate and reliable temperature monitoring. It features an advanced internal sensor with a measuring range of -30°C to +70°C (-22°F to +158°F) and a memory capacity of up to 14,400 readings. This device is suitable for various applications requiring precise temperature recording, such as cold chain logistics, food storage, pharmaceutical monitoring, and laboratory use.
Nastavitev
Before initial use, the Frigga M2D data logger requires configuration using the complimentary Frigga Data Software. This software is compatible with Windows and Mac OS and allows you to set logging conditions such as delayed start, logging intervals, temperature units (Fahrenheit/Celsius), time zones, and temperature alarm ranges. The software can be downloaded from the official Frigga webmesto.
Important Configuration Notes:
- Ensure the 'Stop' button function is set to 'Allow' in the software settings.
- Ensure the 'Restart' option is set to 'Allow' for subsequent uses.
- The device is factory-set with a 30-minutni zamik začetka. During this delay, the logger will not immediately begin recording after pressing the start button. An hourglass icon will appear on the display, and the blue light will flash every 10 seconds. This delay allows the logger to acclimate to its environment before recording begins, preventing initial out-of-range readings.

Image: The Frigga M2D data logger connected to a computer, illustrating the software interface used for setting various parameters like temperature units, alarm ranges, and logging intervals.
Navodila za uporabo
Starting and Stopping Recording
- Za začetek recording, press the 'START/STATUS' button. The device will begin logging data after any configured start delay.
- Za stop recording, press the 'STOP' button.
ViewPodatki v realnem času
The integrated LCD screen displays real-time temperature readings, alarm status, and statistical information. You can switch between Fahrenheit and Celsius units directly on the device.
LED indikator alarma
The device features an LED alarm function that activates if the recorded temperature exceeds or falls below the pre-set thresholds configured in the software.
Pridobivanje podatkov
After recording is complete, plug the USB connector of the data logger into any computer. The device will automatically generate PDF and CSV reports, which can be accessed directly without requiring additional software or driver installation. The PDF report provides a summary and graph, while the CSV file contains raw data points.

Image: The Frigga M2D data logger connected via USB to a computer, demonstrating its ability to auto-generate PDF and CSV reports for data analysis.
Vzdrževanje
Zamenjava baterije
The Frigga M2D uses a CR2450 battery. When the battery indicator shows low power, replace the battery to ensure continuous operation and data integrity. Follow these steps:
- Poiščite predal za baterije na zadnji strani naprave.
- Use a coin or a flat tool to twist the battery cover counter-clockwise to open.
- Odstranite staro baterijo CR2450.
- Vstavite novo baterijo CR2450 s pozitivno stranjo (+) obrnjeno navzgor.
- Replace the battery cover and twist clockwise to secure.

Image: A step-by-step visual guide demonstrating the process of replacing the CR2450 battery in the Frigga M2D data logger.
Čiščenje
Napravo obrišite z mehko krpo,amp cloth. Do not use abrasive cleaners or immerse the device in water, despite its IP65 rating, to prevent damage to the USB port or internal components.
Odpravljanje težav
- Device does not start recording immediately: The device has a factory-set 30-minute start delay. Wait for this period to elapse. An hourglass icon and a flashing blue light indicate the delay is active. Ensure the 'Start Delay' setting in the Frigga Data Software is configured as desired.
- Software issues (e.g., blank window, crashes): If the Frigga Data Software does not open correctly (e.g., a blank window appears), try closing the process via Task Manager and restarting the software. Ensure your operating system (Windows/Mac OS) is updated and meets the software's requirements.
- Inaccurate or unexpected readings: Ensure the device has acclimated to the environment before starting a recording session. Verify that the temperature alarm ranges and logging intervals are correctly set in the Frigga Data Software.
- Računalnik ne prepozna naprave: Ensure the USB connection is secure. Try plugging the device into a different USB port or another computer.
- Kratka življenjska doba baterije: The initial battery might have a reduced charge. Replace the CR2450 battery with a new one if you experience unusually short operational periods.
Če imate vztrajne težave, si oglejte razdelek za podporo, kjer so kontaktni podatki.
Specifikacije
| Model | M2D |
| Merilno območje | -30 °C do +70 °C (-22 °F do +158 °F) |
| Natančnost | ±0.5°C (-20°C to +40°C); ±1°C for other ranges |
| Zmogljivost pomnilnika | 14,400 branj |
| Baterija | CR2450 (zamenljiv) |
| Povezljivost | USB |
| Vrsta zaslona | Digitalni LCD |
| Vodoodporen razred | IP65 |
| Mere (pribl.) | 11.2 cm (4.41 in) Height x 4.7 cm (1.85 in) Width |
| Posebne lastnosti | Alarm, Fast Reading System, High Accuracy, Low Battery Indicator |

Image: A visual representation of the Frigga M2D data logger, highlighting its physical dimensions and core technical specifications.
Garancija in podpora
The Frigga M2D data logger comes with a warranty. Details regarding the warranty period and terms can be found on the warranty card included with your product. You can also scan the QR code on the warranty card for access to the user manual and additional resources.
For technical assistance or any inquiries, 24/7 technician support is available via email. Please refer to the contact information provided on the warranty card or the official Frigga webmesto.
A calibration certificate can be exported by connecting the device to your computer.