1. Uvod
This manual provides essential information for the safe and effective operation of your MiBoce Digital Insulation Resistance Tester. This intelligent instrument utilizes microcomputer technology, large-scale integrated circuits, and digital circuits to provide powerful measurement and data processing capabilities. It is designed for measuring insulation resistance, voltage, and low resistance, making it suitable for on-site power equipment and power supply line inspection and maintenance.
2. Varnostne informacije
Please read and understand all safety warnings and operating instructions before using this instrument. Failure to follow these instructions may result in electric shock, fire, or damage to the instrument.
- Visoka glasnosttage Opozorilo: This device generates high voltage during insulation resistance tests. Always ensure proper connections and never touch test leads or the circuit under test during operation.
- Varnost baterije: Do not mix old and new batteries. Ensure correct battery polarity during installation. Do not operate the instrument with the battery compartment open.
- Okoljski pogoji: Use the instrument within specified environmental limits (temperature, humidity). Avoid using in wet or explosive atmospheres.
- Pregled: Before each use, inspect the instrument and test leads for any damage. Do not use if damaged.
- Kvalificirano osebje: Only qualified personnel should perform electrical measurements and maintenance.
3. Značilnosti izdelka
The MiBoce Digital Insulation Resistance Tester offers a range of features designed for efficient and accurate measurements:
- Test Voltages: Selectable test voltages of 100V, 250V, 500V, and 1000V.
- Samodejni obseg: Automatic ranging for insulation resistance measurements.
- Shranjevanje podatkov: Capability to store measurement data for later review.
- Comparison Measurement: Allows for comparison of current readings against a set reference.
- Polarization Index (PI) Measurement: Automatically calculates the PI ratio based on timed measurements.
- Dielectric Absorption Ratio (DAR) Measurement: Automatically calculates the DAR ratio.
- LCD osvetlitev ozadja: Zagotavlja jasno vidljivost odčitkov v različnih svetlobnih pogojih.
- Analogni stolpčni graf: A 30-segment analog bar graph provides a quick visual indication of resistance trends.
- Over-range Warning: Alerts the user when measurements exceed the instrument's range.
- Visoka glasnosttage Indikacija: Visual indicator when high voltage is present at the test leads.
- Timing Measurement: Supports timed insulation resistance tests, typically around 30 minutes.
- Samodejni izklop: Varčuje z baterijo, saj se po določenem času neaktivnosti samodejno izklopi.
- Indikacija prazne baterije: Opozori uporabnika, ko je treba zamenjati baterije.
4. Nastavitev
4.1 Namestitev baterije
The instrument requires 8 x 1.5V LR14 batteries (not included). To install or replace batteries:
- Ensure the instrument is powered off and all test leads are disconnected.
- Poiščite predal za baterije na zadnji strani enote.
- Odprite pokrov prostora za baterije.
- Insert 8 LR14 batteries, paying close attention to the correct polarity (+/-) as indicated inside the compartment.
- Trdno zaprite pokrov prostora za baterije.

Opis slike: This image shows the underside of the MiBoce Digital Insulation Resistance Tester with the battery compartment cover removed, revealing the slots for eight LR14 batteries. This illustrates the correct location for battery installation.
4.2 Priključitev merilnih vodnikov
The instrument comes with standard accessories including two single-plug test leads with clips (black and green) and one two-plug test lead with a clip (red). Connect the test leads to the appropriate terminals on the front panel:
- V Terminal: Za voltage meritve.
- EARTH Terminal: Ground connection for insulation resistance tests.
- COM terminal: Common terminal for various measurements.
- LINE A Terminal: Line connection for insulation resistance tests.

Opis slike: Ta slika ponuja podroben view of the input terminals (V, EARTH, COM, LINE A) on the top panel of the tester, crucial for connecting test leads. It also shows the side panel with a trap door, adapter jack button, and USB insertion hole, indicating connectivity options.
5. Navodila za uporabo
5.1 Vklop/izklop
Pritisnite VKLOP/IZKLOP button to power the instrument on or off.
5.2 Prikaz končanview
The large LCD provides clear readings and various indicators. The backlight can be activated for improved visibility.

Opis slike: This image focuses on the LCD display of the tester, showing measurement data, function symbols, and a clear digital readout. The display indicates a 100V test voltage, resistance in MΩ, and a timer, demonstrating the clarity provided by the backlight.
5.3 Merjenje izolacijskega upora
- Connect the test leads to the circuit under test (e.g., LINE A and EARTH terminals).
- Izberite želeno voltage (100V, 250V, 500V, or 1000V) using the appropriate buttons.
- Pritisnite TEST button to initiate the measurement. The high voltagIndikator bo zasvetil.
- Read the insulation resistance value on the LCD. The analog bar graph provides a visual trend.
- Pritisnite E-STOP or release the TEST button to stop the measurement and discharge the circuit.
5.4 PI and DAR Measurement
The instrument can automatically calculate Polarization Index (PI) and Dielectric Absorption Ratio (DAR). Refer to the on-screen prompts and specific buttons for these functions.
5.5 Shranjevanje in priklic podatkov
Uporabite SHRANI button to store current measurement data. The NALOŽITEV button allows you to recall previously saved data.
5.6 DC/AC zvtage Merjenje
Connect the test leads to the circuit. Select the DCV/ACV mode to measure DC or AC voltage up to 1000V DC or 750V AC.
6. Vzdrževanje
6.1 Čiščenje
Wipe the instrument's casing z mehkim, damp krpo. Ne uporabljajte abrazivnih čistil ali topil. Pazite, da v instrument ne vdre vlaga.
6.2 Zamenjava baterije
When the low battery indicator appears on the display, replace all 8 LR14 batteries as described in Section 4.1. Remove batteries if the instrument will not be used for an extended period.
6.3 Shranjevanje
Store the instrument in a cool, dry place, away from direct sunlight and extreme temperatures. Keep it in its original packaging or a protective case to prevent damage.
7. Odpravljanje težav
- Instrument se ne vklopi: Preverite namestitev baterij in se prepričajte, da so nove.
- Ni odčitka ali odčitavanje nestabilno: Ensure test leads are securely connected to both the instrument and the circuit under test. Check for damaged leads.
- "OL" or Over-range indication: The measured value exceeds the instrument's maximum range for the selected test voltage. This may indicate an open circuit or extremely high resistance.
- Indikator prazne baterije: Replace all 8 LR14 batteries.
- Visoka glasnosttage warning remains after test: Ensure the circuit has been fully discharged. Do not touch test leads until the high voltage indicator is off.
8. Specifikacije

Opis slike: This image displays the MiBoce Digital Insulation Resistance Tester from an angled front view, highlighting its physical dimensions: 202mm in height, 155mm in width, and 94mm in depth. This provides a clear visual reference for the product's size.
| Parameter | Specifikacija |
|---|---|
| Test izolacijske upornosti Voltages | 100V, 250V, 500V, 1000V (0%~20% tolerance) |
| Insulation Resistance Range (100V) | 0.1MΩ ~ 500MΩ; ±(3%+5) |
| Insulation Resistance Range (250V) | 0.5MΩ ~ 2GΩ; ±(3%+5) |
| Insulation Resistance Range (500V) | 1MΩ ~ 4GΩ; ±(3%+5) |
| Insulation Resistance Range (1000V) | 2MΩ ~ 10GΩ; ±(3%+5) |
| DC Voltage (V) | Up to 1000V; ±(2%+3) |
| AC Voltage (V) | Up to 750V; ±(2%+3) |
| Tok kratkega stika | <2mA |
| Low Resistance (Ω) | 0.1Ω ~ 999.9Ω; ±(1%+3) |
| Napajalnik | 8 x 1.5V LR14 Batteries |
| Velikost zaslona | 123 mm x 58 mm |
| Barva izdelka | Rdeča in siva |
| Neto teža izdelka | 1.2 kg |
| Dimenzije izdelka | 202 mm x 155 mm x 94 mm |
| Standardni dodatki | 2pcs one-plug test lead to one clip (black and green), 1pcs two-plug test lead to one clip (red) |
9. Garancija in podpora
Specific warranty information and customer support contact details are not provided within this manual. Please refer to the product packaging or the seller's webza garancijske pogoje in možnosti podpore.