MAXUS D214

MAXUS Digital Milligram Scale

Model: D214 - User Instruction Manual

Uvod

Thank you for choosing the MAXUS Digital Milligram Scale. This precision instrument is designed for highly accurate measurements up to 50g with a readability of 0.001g, making it ideal for a wide range of applications including powders, jewelry, and other delicate items. Please read this manual thoroughly before operation to ensure proper use and to maximize the lifespan of your scale.

Vsebina paketa

Prosimo, preverite, ali so v vašem paketu priloženi vsi spodaj navedeni artikli:

  • MAXUS Digital Milligram Scale (Model D214)
  • 50 g kalibracijske uteži
  • Powder Trickler
  • Posoda za tehtanje
  • Pinceta
  • zajemalka
  • USB kabel
  • AAA baterije (2)
MAXUS Digital Milligram Scale and included accessories

Image: MAXUS Digital Milligram Scale with all included accessories, including the scale, calibration weight, powder trickler, weighing pan, tweezers, scoop, USB cable, and batteries.

Nastavitev

  1. Napajanje: The scale can be powered by 2 AAA batteries (included) or via the included USB cable. Insert batteries into the compartment on the underside of the scale, or connect the USB cable to a power source.
  2. Umestitev: Place the scale on a stable, level, and vibration-free surface. Avoid areas with drafts, extreme temperature changes, or direct sunlight, as these can affect accuracy.
  3. Začetni vklop: Pritisnite VKLOP/IZKLOP button to turn on the scale. The display will show "88888" and then "0.000g".
MAXUS Digital Milligram Scale power options

Image: Illustration showing the two power supply options for the MAXUS Digital Milligram Scale: Battery Mode and USB Mode.

Navodila za uporabo

  1. Vklop/izklop: Pritisnite VKLOP/IZKLOP button to turn the scale on or off.
  2. Pretvorba enote: Pritisnite M (Mode) button to cycle through the available weighing units: g (grams), oz (ounces), ozt (troy ounces), dwt (pennyweights), ct (carats), and gn (grains).
    MAXUS Digital Milligram Scale unit conversion modes

    Image: Diagram illustrating the six different weighing units available on the MAXUS Digital Milligram Scale: g, oz, ozt, dwt, ct, and gn, changed by pressing the 'M' button.

  3. Funkcija tare: Use the Tare function to subtract the weight of a container, allowing you to measure only the contents.
    1. Turn on the scale and wait for it to stabilize at "0.000g".
    2. Place the empty container on the weighing pan. The display will show the container's weight.
    3. Pritisnite T (Tare) button. The display will reset to "0.000g".
    4. Add the item(s) to be weighed into the container. The display will show the net weight of the items.
    MAXUS Digital Milligram Scale tare function steps

    Image: Four-step visual guide demonstrating the Tare Function: 1. Open product and press ON/OFF. 2. Put item on weighing pan. 3. Press 'T' key to reset reading to zero. 4. Get the weight of the added substance.

  4. Tehtanje: Place the item directly onto the weighing pan or into a tared container. Wait for the reading to stabilize.
  5. Samodejni izklop: The scale features an auto shut-off function to conserve battery power. It will automatically turn off after a period of inactivity.

Umerjanje

Regular calibration ensures the highest accuracy of your scale. Use the included 50g calibration weight for this process.

  1. Ensure the scale is on a stable surface and turned off.
  2. Pritisnite in držite M (Mode) button. While holding, press the VKLOP/IZKLOP button. Release both buttons when the display shows "CAL".
  3. The display will then flash "50.000g".
  4. Carefully place the 50g calibration weight in the center of the weighing pan.
  5. Wait for a few seconds. The display will show "PASS" and then return to "0.000g". Calibration is complete.
MAXUS Digital Milligram Scale calibration steps

Image: Four-step visual guide demonstrating the Calibration process: 1. Press and hold 'M' key until display reads "CAL". 2. Press 'M' key again, display shows "50g". 3. Place 50g weight on platform. 4. LCD displays "PASS".

Vzdrževanje

  • Čiščenje: Obrišite tehtnico z mehko damp krpo. Ne uporabljajte abrazivnih čistil in tehtnice ne potapljajte v vodo.
  • Shranjevanje: Store the scale in a clean, dry environment away from extreme temperatures, humidity, and vibrations. Keep the protective cover closed when not in use.
  • Zamenjava baterije: If the battery indicator appears low or the scale does not power on, replace the AAA batteries.

Odpravljanje težav

TežavaMožen vzrokrešitev
Tehtnica se ne vklopiLow batteries; Incorrect battery installation; No USB power.Replace batteries; Check battery polarity; Ensure USB cable is properly connected to a power source.
Netočne ali nestabilne meritveUnstable surface; Drafts or vibrations; Needs calibration; Overload; Temperature fluctuations.Place on a firm, level surface; Avoid drafts; Perform calibration; Do not exceed 50g capacity; Allow scale to acclimate to room temperature.
Prikazano je "O-Ld" ali "EEEE"Preobremenitev.Remove the item immediately. The scale's capacity is 50g.
"UNST" displayedNestabilno okolje.Ensure the scale is on a stable, vibration-free surface.

Specifikacije

  • Znamka: MAXUS
  • Model: D214
  • Kapaciteta: 50 g
  • Natančnost: 0.001 g
  • Tehtalne enote: g, oz, ozt, dwt, ct, gn
  • Vrsta zaslona: LCD-zaslon z osvetlitvijo ozadja
  • Posebne lastnosti: Samodejni izklop, funkcija tare
  • Možnosti napajanja: 2 x AAA Batteries or USB Cable
  • Material: Akrilonitril butadien stiren, nerjaveče jeklo
  • Dimenzije izdelka: 5.43 x 3.15 x 0.98 palcev
  • Teža artikla: 11.2 unč
MAXUS Digital Milligram Scale dimensions

Image: Diagram showing the dimensions of the MAXUS Digital Milligram Scale: 5.43 inches length, 3.15 inches width, and 0.98 inches height.

Garancija in podpora

For warranty information, technical support, or any inquiries regarding your MAXUS Digital Milligram Scale, please refer to the contact information provided on the product packaging or visit the official MAXUS webspletnega mesta. Račun shranite kot dokazilo o nakupu za uveljavljanje garancije.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.