RCYAGO DO9100
Merilnik raztopljenega kisika s tekočino za polnjenje elektrod
Brand: RCYAGO | Model: DO9100
1. Uvod
The RCYAGO Dissolved Oxygen Meter is a high-precision instrument designed for accurate measurement of dissolved oxygen (DO) and temperature in various water bodies. It utilizes a polarographic sensor with a thermistor for automatic temperature compensation, ensuring reliable readings across different conditions. This manual provides comprehensive instructions for setup, operation, and maintenance of your DO meter.

Figure 1: RCYAGO Dissolved Oxygen Meter and Probe
2. Kaj je v škatli
Prosimo, preverite vsebino paketa in se prepričajte, da so prisotni vsi elementi:
- 1 x RCYAGO Dissolved Oxygen Meter (Model: DO9100)
- 1 x Electrode Probe
- 6 x Probe Protective Covers (Membrane Sleeves)
- 1 x Electrode Filling Solution
- 1 x Anaerobic Water Powder (for calibration)
- 1 x uporabniški priročnik
3. Nastavitev
3.1 Namestitev baterije
The meter requires 3 x 1.5V button batteries. These are typically pre-installed or included in the top cap of the device. Ensure they are correctly oriented before use.
3.2 Priključitev elektrod
Connect the electrode probe to the main meter unit by screwing the metal knob securely. Ensure a firm connection to prevent inaccurate readings.

Figure 2: Meter Components and Dimensions
3.3 Filling the Probe Protective Cover
The probe protective cover contains a permeable membrane. Before first use, or when replacing the membrane sleeve, fill it with the provided electrode filling solution. This solution is crucial for the sensor's proper function. Do not press on the membrane itself.

Figure 3: Highly Sensitive Probe and Protective Covers
4. Delovanje
4.1 Vklop/izklop in prikaz
Pritisnite gumb za vklop (⏻) to turn the meter on. The digital screen will illuminate with a backlight, displaying DO readings and temperature. The backlight improves readability in low-light conditions. The device features an auto-power-off function after 8 minutes of inactivity to conserve battery life.

Figure 4: HD Backlight Display and Battery Compartment
4.2 Načini merjenja
The meter can display dissolved oxygen in mg/L (parts per million, ppm) or as a saturation percentage (%). Press the NAČIN/KAL button to toggle between these modes.
4.3 Temperaturna kompenzacija
The meter features automatic temperature compensation (ATC) within the range of 0-50°C (32-122°F). This ensures that dissolved oxygen readings are accurate despite variations in water temperature, as oxygen solubility is affected by temperature.

Figure 5: Automatic Temperature Compensation in action
4.4 Kalibracija
Regular calibration is essential for accurate measurements. The meter supports two-point calibration: 100% air calibration and 0% anaerobic water calibration.
4.4.1 100% Air Calibration
- Ensure the probe is exposed to air.
- Dolgo pritisnite NAČIN/KAL gumb za 5 sekundi.
- Release the button. The screen will flash 3 times at 100%. Calibration is complete.
4.4.2 0% Anaerobic Water Calibration
- Prepare anaerobic water: Pour 1 packet of anaerobic powder into the attached calibration bottle, add 30ml of purified water or tap water, and wait 5-10 minutes for the powder to fully dissolve.
- Submerge the electrode in the anaerobic water.
- Pritisnite NAČIN/KAL key to switch to mg/L mode.
- Wait for the value to drop to 0.
- Dolgo pritisnite NAČIN/KAL gumb za 5 sekundi.
- Release the button. The screen will flash 3 times at 0%. Calibration is complete.

Figure 6: Calibration Procedures
4.5 Merjenje
Once calibrated, submerge the electrode probe into the water sample you wish to measure. The meter will display the dissolved oxygen and temperature readings. Allow a few moments for the readings to stabilize.

Figure 7: Measuring Dissolved Oxygen in a Pond
4.6 Preklapljanje temperaturnih enot
To switch between Celsius (°C) and Fahrenheit (°F), long press the ZADRŽI/TEMP gumb.
4.7 Understanding Oxygen Solubility
Oxygen solubility in water is influenced by temperature, atmospheric pressure, and salinity. The provided table illustrates the saturation concentration of oxygen at different temperatures:

Figure 8: Oxygen Saturation Concentration at Different Temperatures
5. Vzdrževanje
5.1 Probe Protective Cover (Membrane Sleeve)
The film inside the protective cover cannot be taken apart. If test values become inaccurate even after calibration, the membrane sleeve needs to be replaced. The meter comes with 6 spare covers.
5.2 Čiščenje
The meter unit and probe should be cleaned with care. The product care instructions specify 'Hand Wash Only'. Avoid harsh chemicals or abrasive materials.
5.3 Shranjevanje
When not in use, ensure the probe protective cover is properly filled with solution and securely attached to protect the sensor. Store the meter in a cool, dry place.
6. Odpravljanje težav
- Netočni odčitki: Ensure the electrode is properly connected and the membrane sleeve is filled with the electrode filling fluid. Perform calibration regularly (both air and anaerobic water calibration) as described in Section 4.4.
- Težave s prikazom: Check battery levels. If the display is dim or erratic, replace the batteries.
- Brez branja: Verify the electrode is fully submerged in the sample and that the connection to the meter is secure.
- Napaka kalibracije: Ensure the anaerobic water powder is fully dissolved for 0% calibration. For air calibration, ensure the probe is exposed to stable ambient air. If issues persist, the membrane sleeve might need replacement.
7. Specifikacije
| Funkcija | Podrobnost |
|---|---|
| Dimenzije izdelka | 6.1 x 0.01 x 0.01 palcev |
| Številka modela artikla | DO9100 |
| Teža | 12.56 unč |
| Znamka | RCYAGO |
| Posebna funkcija | Visoka natančnost |
| barva | Črna |
| Zunanji material | Plastika |
| Vrsta zaslona | Digitalni z osvetlitvijo ozadja |
| Dissolved Oxygen Range | 0.0-40.0 mg/L |
| Dissolved Oxygen Saturation | 0.0-300.0 % |
| Temperaturno območje | 0.0-40.0 °C (32.0-104.0 °F) |
| Samodejna temperaturna kompenzacija | 0-50°C |
| Vir napajanja | 3 x 1.5V Button Batteries |
| Samodejni izklop | 8 minut |
8. Videoposnetki izdelkov
8.1 Digital Dissolved Oxygen Meter - Basic Operation & Calibration
Ta videoposnetek ponuja hiter pregledview of the RCYAGO Dissolved Oxygen Meter, demonstrating how to power it on, install the electrode, and perform both air and oxygen-free water calibrations. It highlights the key steps for initial setup and ensuring accurate readings.
8.2 Portable Dissolved Oxygen Meter - Detailed Review & Uporaba
This comprehensive video offers a detailed review of the portable dissolved oxygen meter. It covers unboxing, battery installation, electrode setup, and practical demonstrations of measuring dissolved oxygen in different water samples (tap water vs. boiled water). The video also discusses the included accessories like membrane caps and electrolyte solution, and explains the temperature conversion chart.
9. Garancija in podpora
For warranty information, technical support, or to purchase replacement parts (such as probe protective covers or electrode filling solution), please contact RCYAGO customer service. Refer to the product packaging or the official RCYAGO webspletno mesto za kontaktne podatke.